203
Update
Appendix I Firmware Up
g
rade
UPGRADING USB BRIDGE FIRMWARE
1) Connect the USB cable between the scanner and your computer.
If you are using USB Virtual COM for the first time, you must install its driver from
the CD-ROM.
2) Read the following barcodes in sequence to configure the scanner to use USB Virtual
COM as download interface.
Enter Setup
Activate USB
Virtual COM
Update
3) Read the following barcodes in sequence for the scanner to enter the download mode.
The scanner will respond with beeps to indicate it is ready for downloading.
Enter Setup
Download USB Bridge
Firmware
4) Run the download utility “ProgLoad.exe” on your computer.
Kernel Program
User Program
K170xBridge_V*.shx STD170xBridge_V*.shx